1 | Make sure that the currently connected transducer(s) are shown as "tabs" at the bottom of the EK80 presentation. |
2 | Open the Setup menu. |
3 | On the Setup menu, select Installation. Observe that the Installation dialog box opens. This dialog box contains a number of pages selected from the menu on the left side.
|
4 | On the left side, select Transducer Installation. On the Installed Transducers list, select one of the transducers. Observe that the Transducer page opens with all settings unavailable. This is a safety precaution to prevent unintentional changes to the transducer settings. To make any changes, you must select Edit.
|
5 | Make sure that each transducer has been installed with all settings defined. The physical location of the transducer is important for the echo data accuracy. The offset value for "Z" defines the vertical location (the depth) of the transducer face.
This is the installation depth of the transducer; the vertical location of the transducer face relative to the water surface. In order to measure correct water depth, the EK80 needs to know the vertical distance between the vessel's water line and the acoustic face of each transducer. The depth of each individual transducer must be defined manually. Enter the depth as a positive number. If the displacement of your vessel changes considerably, you may consider changing this parameter often. For accurate location of the transducer, you need the detailed vessel drawings.
|
6 | On the left side of the Installation dialog box, select Transceiver. Observe that the Transceiver Installation page opens.
|
7 | Make sure that all applicable transceivers and transducers are connected and operational. For each transceiver, this is indicated by the green label with text "Installed".
|
8 | Close the Installation dialog box without making any changes. |