• | The system units have all been installed according to the instructions in the Installation manual. |
• | All cable connections have been made and checked. |
• | All system units have been inspected. |
• | The operational software is available. |
• | The software license is available. |
1 | Do the following preparations.
|
2 | Install the software. |
3 | Turn on each transceiver. |
4 | Make sure that the Processor Unit is connected to the transceiver(s) using the Ethernet cable specified in the Installation manual. If you use more than one transceiver, a high capacity Ethernet switch is required.
Note
It is very important that high-quality Ethernet cables are used. You must use CAT-5E STP (Shielded Twisted Pair) quality or better. Using cables with lower bandwidth capacity will reduce performance.
|
5 | Start the EK80 program. |
6 | On the Software License page, install the software license(s). |
7 | On the Transducer Installation page, install each transducer not automatically installed. |
8 | On the Transceiver Installation page, connect the transducer(s) to the transceiver(s). |
9 | On the Sensor Installation page, set up the interfaces to the navigation sensors. |
10 | In the Environment dialog box, select correct water temperature and salinity. |
11 | Start normal operation.
|