Prerequisites
Neither tools nor instruments are required. The EK80 system is turned on and operates normally.
|
• |
For "slave" operation, a remote system (for example K-Sync or Simrad TU40) must be available to provide trigger pulses. |
|
• |
For "master" operation, a remote hydroacoustic system (sonar, echo sounder) is connected. This remote system must be set up in "slave" mode. |
The external system is connected to a serial communication port on the Processor Unit.
Context
Whenever more than one hydroacoustic system is installed on a vessel, interference may occur. To avoid interference, you have these options:
|
• |
The systems are all connected to a common synchronization system. |
|
• |
One of the acoustic systems is set up as "master", and controls the transmissions on the other systems. |
Procedure
|
1 |
On the Setup menu, select Installation. |
|
2 |
Select Synchronization. |
|
3 |
Choose which serial port you have used to connect the external synchronization system. |
|
4 |
Select Apply and then Close to save all the parameters and close the Installation dialog box. |