Calibration spheres
Kongsberg Maritime can supply a variety of copper and tungsten calibration spheres dedicated for different operational frequencies. Each sphere diameter is selected for minimum temperature dependence.
Frequencies (kHz) Copper Part number
18 63 mm SRT-073512
38 60 mm SRT-073514
70 32.1 mm SRT-073517
120 23 mm SRT-088098
200 13.7 mm SRT-073519
Frequencies (kHz) Tungsten carbide Part number
38 38.1 mm 337897
70 38.1 mm 337897
120 38.1 mm 337897
200 38.1 mm 337897
333 22 mm 337511
- 25 mm 364507
Each tungsten carbide sphere must have a cobalt binder.
Note
Each calibration sphere must be handled with care to avoid any damage to its surface. When not in use, store the sphere in a household soap solution. All suspension lines must be as thin and clean as possible. Limit knots to a minimum, and keep them small. Even knots have strong echoes!