10 Recording processed data
The Record Processed function provided by the EK80 system allows you to save processed data on a selected format. The data files can be copied or moved to an external storage device, or to another computer on the network. You can keep the recorded files for scientific studies, future references or for training purposes. The recording is controlled by the Record Processed function.
Prerequisites
Before you start data recording, make sure that you have defined where to store the files. To define which disks and folders to use to save the data files, use the File Setup page. The File Setup page is located in the Output dialog box. If you wish to save your recorded data on an external hard disk, make sure that it is connected to the computer.
Before you start recording, you must also define which output format(s) you wish to use. To choose which processed data formats to record, select Processed Data Output. The Processed Data Output page is located in the Output dialog box.
Context
You can save and recall echo information using the following methods and formats:
•  Bitmap images (containing the full screen) are saved when you select Screen Capture on the top bar.
•  Use Record RAW on the Operation menu to record raw data.
•  A "history file" is recorded automatically and continuously. When the file is full it will start to overwrite the oldest data, thus creating a "ring buffer".
•  Use Record Processed on the Operation menu to record processed data.
This is only an export format. Processed data files can not be played back on the EK80 system.
Note
The data files will normally become very large. If you wish to record large amounts of data, make sure that you have enough space on your hard disk. Unless your computer is equipped with a very large disk, we recommend that you save the data to an external storage device.
You can record both RAW and processed data using the Record RAW and Record Processed functions. It may be useful to synchronize these two functions to automatically record all data formats simultaneously. To synchronize the recording functions, open the Record RAW button, and set Synchronize to On. The Record Processed button is then inhibited, and you can start and stop all recording by means of the Record RAW button.
Procedure
1 Open the Operation menu.
2 Make sure that the Record Processed function is available.
Image
If the Record Processed function is unavailable, it is most likely because you have forgotten to specify an output format.
a On the Operation menu, select Output.
b On the left side of the Output dialog box, select Processed Data Output to open the page.
c Select New in the Installed Outputs area.
d Select an output type from the list.
e Select the settings for the chosen output format.
f Select Add to save the output format.
g At the bottom of the Output dialog box, select OK to save the chosen parameters and close it.
3 To start data recording, open the Record Processed button, and select On.
Alternatively, simply select the red circle on the right side of the button.
4 To stop recording, open the Record Processed button, and select Off.
Alternatively, select the black rectangle on the left side of the button to stop the recording.