| 1 | Disconnect all transducers by removing the transducer cables from the transceiver. |
| 2 | Replace the transducers with a dummyload. |
| 3 | Record raw data.
|
| 4 | Open the Setup menu. |
| 5 | Open the Diagnostics dialog box from the Setup menu.![]() |
| 6 | Select Noise to open the page. |
| 7 | Read and record the Noise Estimate value for 10 pings. The value will vary from ping to ping.
|
| 8 | Calculate the average noise estimate from the values of the10 ping. |
| 9 | Enter this value in the table. |